Technical Pitfalls in Optic Nerve Sheath Diameter POCUS: A Practical Checklist for Image Optimization

Ricardo Bañuelos-Huerta, MD MSc MPH1*; Dibanhi S. Ruvalcaba-Trujillo, MD2; Jorge A. Lara-Arellano, MD3; Gilver Sebastián-Vázquez, MD4; Carlos Estrada-Valencia, MD5; Wendy J. Ramirez-Salvador, MD6; Daniel de J. Vásquez-Hernández, MD7; Montserrat Ramírez-Barranco, SSP8; Eduardo Y. Aguilar-Lazcano, SSP8; Wendy Sartillo-Guerra, SSP8

(1) Department of Emergency Medicine, Hospital General Tlaxcala “Lic. Anselmo Cervantes Hernández”, Tlaxcala, Mexico

(2) Department of Emergency Medicine, Hospital General Zona No. 30, Instituto Mexicano del Seguro Social, Mexico City, Mexico

(3) Department of Emergency Medicine, Hospital General Zona No. 3, Instituto Mexicano del Seguro Social, Sinaloa, Mexico

(4) Department of Emergency Medicine, Hospital Regional de Alta Especialidad “Bicentenario de la Independencia”, Instituto de Seguridad y Servicios Sociales de los Trabajadores del Estado, Estado de Mexico, Mexico

(5) Department of Emergency Medicine, Hospital General de Zona No. 1 “Dr. Luis Ernesto Miramontes Cárdenas”, Instituto Mexicano del Seguro Social, Nayarit, Mexico

(6) Department of Emergency Medicine, Hospital General de Zona No. 1, Instituto Mexicano del Seguro Social, Chiapas, Mexico

(7) Department of Emergency Medicine, Hospital General de Zona No. 35, Instituto Mexicano del Seguro Social, Puebla, Mexico

(8) Department of Research, Hospital General de Tlaxcala “Lic. Anselmo Cervantes Hernández”, Tlaxcala, Mexico

*Corresponding Author: Dr. Ricardo Bañuelos Huerta (email: dr.banuelos.urgencias@gmail.com)


Download article PDF – POCUS Journal 2026;11(2):143-148

DOI: https://doi.org/10.24908/pocusj.v11i02.20719


Abstract

Background: Point of care ultrasound (POCUS) measurement of the optic nerve sheath diameter (ONSD) has been established as a non-invasive method for estimating intracranial pressure in critically ill patients. However, the accuracy of the technique depends strongly on proper image acquisition. Errors in system optimization may result in inaccurate measurements and increased inter-operator variability. Objective: Our objective was to describe the most common technical errors in ONSD measurement and to propose a checklist to optimize image acquisition. Methods: A narrative review of the literature and an analysis of technical errors observed during POCUS training in critical care settings were completed. Emphasis was on ultrasound system parameters and image acquisition. Results: Six key technical elements were identified to optimize ONSD measurement. They included preset selection, depth adjustment, focal zone positioning, transducer frequency, gain optimization, and identification of the central retinal artery and vein. Conclusion: Standardization of ultrasound system settings may improve image quality and reduce measurement errors in ONSD, increasing reproducibility in emergency and critical care settings.

Introduction

Intracranial hypertension (ICH) is a frequent complication in various acute neurological conditions, including traumatic brain injury, intracranial hemorrhage, neurologic infections, and cerebral edema. Invasive intracranial pressure monitoring remains the diagnostic gold standard; however, its availability is limited in many clinical settings.

Over the past two decades, POCUS measurement of the optic nerve sheath diameter (ONSD) has become a widely used non-invasive method for estimating increased intracranial pressure. Several studies have demonstrated a significant correlation between increased ONSD and elevated intracranial pressure [1,2]. Hansen and Helmke were the first to describe the relationship between the perioptic subarachnoid space and intracranial pressure changes using ultrasound [3,4]. Subsequent clinical studies confirmed that ONSD enlargement correlates with ICH in patients with traumatic brain injury and other neurological conditions [5–7].

Soldatos et al. reported that an ONSD greater than 5.7 mm has high specificity for predicting intracranial pressure above 20 mmHg [8]. Similarly, Moretti and Pizzi demonstrated a strong correlation between ONSD and invasively measured intracranial pressure [9]. A meta-analysis by Dubourg et al. confirmed a sensitivity of approximately 90% and specificity of 85% for detecting ICH using optic nerve ultrasound [1].

Despite its apparent technical simplicity, ONSD measurement is subject to significant errors related to image acquisition. A considerable proportion of inaccurate measurements results from suboptimal ultrasound system settings rather than clinical interpretation [10] (Figure 1).

Figure 1. Image obtained using a superficial musculoskeletal (MSK) preset, where the optic nerve is not adequately visualized. The optic nerve sheath cannot be identified, and artifacts related to excessive gain are observed.

In clinical practice, many operators perform ONSD measurements without adequately optimizing image parameters, leading to systematic errors such as overestimation of sheath diameter, misidentification of the optic nerve, or measurements obtained outside the true central axis of the nerve.

Although ONSD measurement has become a valuable non-invasive tool for estimating intracranial pressure, there is still no universal consensus regarding the optimal measurement technique. The CLOSED protocol summarizes essential steps for accurate and safe measurement (Colour Doppler, Low power examination, Optic disc clarity, Safety, Elevated frequency, Dual measurements), allowing for more consistent and reproducible results. However, the limited availability of standardized comparative imaging may contribute to inter-observer variability [11].

Therefore, in order to standardize the technique and improve measurement reproducibility, we propose a technical checklist based on six key steps prior to measurement, focused on image quality optimization and accurate anatomical identification.

Anatomical and Embryological Basis

The optic nerve is a direct extension of the central nervous system that originates from the diencephalon during embryonic development. Unlike peripheral nerves, it is myelinated by oligodendrocytes and is surrounded by the meninges: dura mater, arachnoid mater, and pia mater [12].

Between the arachnoid and pia mater lies the subarachnoid space, which contains cerebrospinal fluid and is continuous with the intracranial subarachnoid space. Due to this anatomical continuity, increases in intracranial pressure are transmitted to the optic nerve sheath, resulting in its distension.

Anatomical and experimental studies have demonstrated that maximal distensibility of the sheath occurs approximately 3 mm posterior to the globe, which is why this point is used as the reference for ONSD measurement [3–4].

Relevant Ultrasound Physics Principles

Ultrasound image quality depends on several technical parameters, including transducer frequency, imaging depth, focal zone positioning, and gain settings.

High-frequency linear transducers (7–15 MHz) provide superior spatial resolution for superficial structures such as the optic nerve. As frequency increases, axial and lateral resolution improve, although tissue penetration decreases [13–15].

Gain controls image brightness by amplifying returning echoes. Excessive gain may produce overexposure artifacts or “blooming artifact,” making it difficult to accurately delineate the anatomical borders of the optic nerve sheath.

Additionally, the vitreous humor behaves as a low-attenuation fluid medium, resulting in the phenomenon known as posterior acoustic enhancement. This artifact causes structures located posterior to the globe to appear artificially brighter, potentially interfering with accurate delineation of the optic nerve [14–15].

Technical Checklist Prior to ONSD Measurement

Based on available literature and clinical experience in ultrasound training in critical care, we propose a technical checklist before ONSD measurement (Table 1).

Table 1. Technical checklist for standardized optic nerve sheath diameter (ONSD) ultrasound acquisition.
Checklist before measuring optic nerve sheath diameter (ONSD)
✓1. Select ocular preset (if it is available) or small parts
✓2. Adjust depth ( ≈ 3-4 cms)
✓3. Use high-frequency linear probe
✓4. Place focal zone at optic nerve
✓5. Reduce global & TGC gain to avoid artifact
✓6. Use color Doppler to identify central retinal artery

1. Preset Selection

An ocular or superficial preset should be selected to optimize resolution for near-field structures. These presets are designed to optimize multiple system parameters, including frequency, dynamic range, image processing, and initial gain settings, thereby improving spatial resolution for superficial structures such as the optic nerve.

Additionally, ocular presets typically incorporate automatic limitations in acoustic output, helping to maintain safe values of the thermal index (TI) and mechanical index (MI) during examination. This is particularly important in ocular ultrasound, as the globe is sensitive to thermal and mechanical effects of ultrasound.

For this reason, international recommendations suggest performing ocular ultrasound while maintaining the TI as low as possible, ideally below 0.5 [13–14,16]. The use of non-specific presets (such as abdominal or vascular) may unnecessarily increase acoustic output and reduce optimization for superficial imaging. Therefore, selecting an ocular preset is a critical step for both image quality and patient safety [12–14].

2. Depth Adjustment

Imaging depth should be limited to approximately 3–4 cm, focusing system resolution on the retrobulbar region where measurement is performed. Excessive depth reduces image quality in the region of interest.

From a physical standpoint, increasing depth forces the ultrasound system to:

 • Reduce line density

 • Decrease effective lateral resolution

 • Distribute pixels over a larger field

This results in reduced spatial resolution of superficial structures, making accurate identification of the optic nerve sheath more difficult [7,12]. Therefore, limiting depth allows concentration of resolution in the relevant anatomical region (Figure 2).

Figure 2. Image with reduced depth, limited to 3–4 cm, allowing improved visualization of the retrobulbar region.

3. Frequency Adjustment

High-frequency linear transducers (7–15 MHz) should be used. Higher frequencies provide:

 • Improved axial resolution

 • Better interface definition

 • Enhanced ability to distinguish small structures

This is crucial for ONSD measurement, as small variations in resolution may lead to clinically significant errors [13–15]. The main limitation of higher frequencies is reduced penetration; however, this is not relevant in superficial structures such as the eye.

4. Focal Zone Adjustment

The focal zone represents the region where the ultrasound beam reaches its narrowest width and highest lateral resolution. It should be positioned at the level of the optic nerve.

When the focal zone is placed outside the region of interest:

 • Beam width increases

 • Lateral resolution decreases

 • Differentiation between sheath and nerve become more difficult

Proper focal positioning significantly improves delineation of the hyperechoic borders of the optic nerve sheath used for measurement.

5. Gain and Time Gain Compensation (TGC) Adjustment

Global gain and Time Gain Compensation (TGC) must be carefully adjusted to optimize contrast between the optic nerve and its sheath. One of the most common technical errors in ONSD measurement is excessive gain, which leads to:

 • Image overexposure

 • Reduced contrast

 • Blooming artifact that artificially enlarges structures

 • Difficulty in delineating retrobulbar anatomy

Additionally, the vitreous humor is a low-attenuation medium, resulting in posterior acoustic enhancement [8]. This artifact occurs because ultrasound waves pass through fluid with minimal energy loss, allowing greater acoustic energy to reach structures posterior to the globe. As a result, retrobulbar structures—including the optic nerve—may appear artificially hyperechoic, making differentiation between the nerve and its sheath more difficult.

To counteract this phenomenon:

 • Reduce global gain

 • Adjust TGC in the region posterior to the globe

 • Optimize contrast to clearly identify sheath borders

In clinical practice, many operators mistakenly interpret increased retrobulbar echogenicity as part of the optic nerve sheath, when it actually represents an artifact caused by posterior acoustic enhancement of the vitreous humor [15–17] (Figures 3 and 4).

Figure 3. (A) Ultrasound image demonstrating posterior acoustic enhancement within the retrobulbar region after gain optimization. (B) Schematic representation of retrobulbar anatomy demonstrating the optic nerve (gray), the external optic nerve sheath (green), internal optic nerve sheath (red) and subarachnoid space (blue).
Figure 4. Ultrasound image demonstrating focal zone positioning at the optic nerve sheath measurement site. The left-sided blue adjustment curve corresponds to the Time Gain Compensation (TGC) settings used to optimize retrobulbar contrast and improve delineation of the optic nerve sheath.

6. Identification of the Central Retinal Artery and Vein

Identification of the central retinal artery and vein using colour Doppler provides an additional anatomical reference that enhances the precision and reliability of the measurement. Locating these vessels ensures that the ONSD measurement is performed in the correct plane, reducing errors associated with oblique imaging planes, which may be difficult to detect using B-mode alone due to acoustic shadowing [11,12]. Measurements obtained outside the central axis tend to overestimate or underestimate ONSD, reducing reproducibility [7] (Figure 5).

Figure 5. (A) Ultrasound image with colour Doppler demonstrating the central retinal artery and vein. (B) Schematic representation of the central retinal vessels and their anatomical relationship with the optic nerve.

Standard Measurement Technique

The patient should be placed in the supine position with the eyelid closed. Sterile gel can be applied over the eyelid to avoid direct pressure on the globe. The linear transducer is then placed in either transverse or longitudinal orientation over the eye. The main ocular structures are identified: anterior chamber, lens, vitreous humor, and posterior globe wall. The optic nerve appears as a hypoechoic tubular structure extending posteriorly from the globe, surrounded by a hyperechoic sheath. ONSD measurement is performed 3 mm posterior to the retinal border by drawing a line perpendicular to the nerve’s longitudinal axis, connecting both internal borders [18]. Although observational cohort studies, such as the study by Pansell et al., suggest that external-to-external border measurements may provide comparable diagnostic accuracy, the international consensus statement by Hirzallah et al. prioritizes the use of the inner edge-to-inner edge technique in clinical practice [19,20]. At least three measurements should be obtained and averaged to improve reproducibility [13].

Conclusion

Ultrasonographic measurement of ONSD is a useful, non-invasive tool for assessing intracranial pressure in critically ill patients. However, its accuracy depends heavily on proper image acquisition. Technical errors related to ultrasound system configuration represent a major source of variability in ONSD measurement and may lead to misinterpretation when imaging parameters are not properly optimized. The implementation of a technical checklist—including appropriate preset selection, depth, focal zone, frequency, gain optimization, and the use of colour Doppler—can significantly improve image quality and reduce measurement errors.

Ocular ultrasound should always be performed following acoustic safety principles, maintaining ultrasound output as low as reasonably achievable. The use of ocular presets and monitoring of TI and MI help ensure safe evaluation of the globe during ONSD measurement. Standardization of these parameters may improve diagnostic accuracy and enhance reproducibility among operators, facilitating integration into clinical practice and POCUS training programs.

Ethics Statement

This article is based on a narrative review of the literature and does not involve human participants, patient data, or identifiable patient information. Therefore, institutional ethics approval and informed consent were not required.

Disclosure Statement

The authors declare no conflicts of interest related to this work.

Funding

This work received no external funding or financial support.

Author Contributions

RBH: conceptualization, methodology, formal analysis, writing – original draft, writing – review & editing, supervision. DSRT: methodology, writing – original draft. JALA: methodology, writing – original draft. GSV: writing – original draft. CEV: writing – original draft. WJRS: methodology, writing – original draft. DJVH: writing – original draft. MRB: formal analysis, writing – original draft. EYAL: methodology. WSG: methodology.

References

1. Dubourg J, Javouhey E, Geeraerts T, Messerer M, Kassai B. Ultrasonography of optic nerve sheath diameter for detection of raised intracranial pressure: a systematic review and meta-analysis. Intensive Care Med. 2011;37:1059–68

2. Koziarz A, Sne N, Kegel F, Nath S, Badhiwala JH, Nassiri F, Mansouri A, Yang K, Zhou Q, Rice T, Faidi S, Passos E, Healey A, Banfield L, Mensour M, Kirkpatrick AW, Nassar A, Fehlings MG, Hawryluk GWJ, Almenawer SA. Bedside optic nerve ultrasonography for diagnosing increased intracranial pressure: a systematic review and meta-analysis. Ann Intern Med. 2019;171:896–905.

3. Hansen HC, Helmke K. The subarachnoid space surrounding the optic nerves: an ultrasound study of the optic nerve sheath. Surg Radiol Anat. 1996;18:323–8.

4. Helmke K, Hansen HC. Fundamentals of transorbital sonographic evaluation of optic nerve sheath expansion under intracranial hypertension. Pediatr Radiol. 1996;26:701–5.

5. Blaivas M, Theodoro D, Sierzenski PR. Elevated intracranial pressure detected by bedside emergency ultrasonography of the optic nerve sheath. Acad Emerg Med. 2003;10:376–81.

6. Geeraerts T, Launey Y, Martin L, Pottecher J, Vigué B, Duranteau J, Benhamou D. Ultrasonography of the optic nerve sheath may be useful for detecting raised intracranial pressure after severe brain injury. Intensive Care Med. 2007;33:1704–11.

7. Chen W, Zhang X, Ye X, Ying P. Diagnostic accuracy of optic nerve sheath diameter on ultrasound for the detection of increased intracranial pressure in patients with traumatic brain injury: a systematic review and meta-analysis. Biomed Rep. 2023;19(6):103. doi:10.3892/br.2023.1685

8. Soldatos T, Karakitsos D, Chatzimichail K, Papathanasiou M, Gouliamos A, Karabinis A. Optic nerve sonography in the diagnostic evaluation of adult brain injury. Crit Care. 2008;12:R67.

9. Moretti R, Pizzi B, Cassini F, Vivaldi N. Reliability of optic nerve ultrasound for the evaluation of patients with spontaneous intracranial hemorrhage. Neurocrit Care. 2009;11:406–10.

10. Ballantyne SA, O’Neill G, Hamilton R, Hollman AS. Observer variation in the sonographic measurement of optic nerve sheath diameter. Eur J Ultrasound. 2002;15:145–9.

11. Aspide R, Bertolini G, Albini Riccioli L, Mazzatenta D, Palandri G, Biasucci DG. A proposal for a new protocol for sonographic assessment of the optic nerve sheath diameter: the CLOSED protocol. Neurocrit Care. 2020;32:327–32.

12. Sadler TW. Langman’s medical embryology. 14th ed. Philadelphia: Wolters Kluwer; 2019.

13. Robba C, Santori G, Czosnyka M, Corradi F, Bragazzi N, Padayachy L, Taccone FS, Citerio G. Optic nerve sheath diameter measured sonographically as non-invasive estimator of intracranial pressure: a systematic review and meta-analysis. Intensive Care Med. 2018;44(8):1284–94. doi:10.1007/s00134-018-5305-7

14. Kremkau FW. Diagnostic ultrasound: principles and instruments. 9th ed. Philadelphia: Elsevier; 2015.

15. Rumack CM, Wilson SR, Charboneau JW. Diagnostic ultrasound. 5th ed. Philadelphia: Elsevier; 2018.

16. American Institute of Ultrasound in Medicine. Ocular ultrasound safety guidelines. J Ultrasound Med. 2013;32:1–3.

17. Di Serafino M, Iacobellis F, Schillirò ML, D’Auria D, Verde F, Grimaldi D, Dell’Aversano Orabona G, Caruso M, Sabatino V, Rinaldo C, Guerriero P, Cantisani V, Vallone G, Romano L. Common and uncommon errors in emergency ultrasound. Diagnostics (Basel). 2022;12:631.

18. Mohammadi A, Haghighi M, Adabi M, Samaeili A, Kayedi M. Optic nerve sheath diameter measurement in healthy South Iranian adults from a cross-sectional ultrasonographic study. Sci Rep. 2025;15:17404.

19. Pansell J, Bell M, Rudberg P, Friman O, Cooray C. Optic nerve sheath diameter in intracranial hypertension: measurement external or internal of the dura mater? J Neuroimaging. 2023;33:58–66.

20. Hirzallah MI, Lochner P, Hafeez MU, Lee AG, Krogias C, Dongarwar D, Hartman ND, Ertl M, Robba C, Malojcic B, Valaikiene J, Sarwal A, Hakimi R, Schlachetzki F; Optic Nerve Sheath Diameter Point-of-Care Ultrasonography Quality Criteria Checklist (ONSD POCUS QCC) Expert Panelists. Optic Nerve Sheath Diameter Point-of-Care Ultrasonography Quality Criteria Checklist: An International Consensus Statement on Optic Nerve Sheath Diameter Imaging and Measurement. Crit Care Med. 2024;52:1543–56.

Posted in .

Leave a Reply

Your email address will not be published. Required fields are marked *